Optimisation of inductive telemetry links in an implantable medical device for a wireless range of 25 cm

B. John, C. Spink, M. Braunschweig, R. Ranjan, D. Schroeder, A. Koops, G. Woldt, G. Adam, W. H. Krautschneider

PublikationBegutachtung

4 Zitate (Scopus)

Abstract

The optimisation of inductive telemetry link for increasing wireless range of a medical implant device is studied in detail. An active implant device using separate wireless power and data links at 2 MHz and 420 kHz has been fabricated and tested. A test setup consisting of RF power transmitter and data reader hardware has been implemented. Measurement results show that the fabricated device achieved a wireless range of 25 cm.

OriginalspracheEnglish
Seiten (von - bis)1315-1316
Seitenumfang2
FachzeitschriftElectronics Letters
Jahrgang54
Ausgabenummer23
DOIs
PublikationsstatusPublished - 15 Nov. 2018